Explantation of Aortic Endograft for Type I Endoleak: 13

A D V E R T I S E M E N T

   
 

A cross sectional view of the explanted endograft is shown. The crimping of the graft posteriorly was the cause of the Type I endoleak, allowing luminal continuity with the aneurysm sac.       

Notes:

Link to this frame from your Personal Thumbnails page? Yes No


 

Explantation of Aortic Endograft for Type I Endoleak: 14

A D V E R T I S E M E N T

   
 

Side by side comparison of the explanted endograft and the preoperative CT, magnified for comparison with the gross specimen. 

Vascular Surgeon: R. Eugene Zierler
Vascular Fellow: Shyam Krishnan
Division of Vascular Surgery
University of Washington, Seattle      